MICRO-EPSILON
X9751109-B060097MSC
7
1.5
Proper Environment
• Protection class Sensor:
IP64 (applies only with sensor cable connected)
• Protection class controller: IP50
• The level of protection does not apply to optical inputs, because when they
become dirty, impairment or failure of the function results.
• Operating temperature:
0 ... 50 °C
• Storage temperature:
-20 ... 70 °C
• Air humidity:
5 - 95 % (non-condensing)
• Ambient pressure:
Atmospheric pressure
• EMC:
According to
EN 55011 Class B /05.200 Emitted interference
EN 61 000-6-2 /08.2002
Immunity to interference
2.
Laser Safety
The scanCONTROL28x0 sensors operate with a semiconductor laser having a
wavelength of 655 nm (visible/red). The laser operation is indicated visually by the LED
on the sensor and on the controller.
When operating the scanCONTROL28x0 sensors, the relevant regulations according to
EN 60825-1 (IEC 60825, Part 1 of 11/2001) and the applicable accident prevention
regulations must be followed.
The housing of the scanCONTROL28x0 optical sensors must only be opened by
authorized persons, (see also Chap. 9). For repair and service, the sensors should
always be returned to the manufacturer.
2.1
Laser Class 2M
Sensors with a maximum laser power up to 15 mW (see Chap. "Technical Data") are
classified in Laser Class 2M (IIM).
Accordingly, the following applies:
With laser equipment of the Class 2M, the eye is not put in danger during random,
short-term exposure to the laser radiation, i.e. exposure duration up to 0.25 s.
A direct glimpse into the beam can be dangerous if the eye-closure reflex is
deliberately suppressed, e.g. during adjustment. Direct viewing into the beam with
optical aids, e.g. a magnifying glass, is dangerous.
Laser equipment of the Class 2M can be employed without further protective measures,
when deliberate viewing into the laser beam or into a beam reflected by mirrors is not
longer than 0.25 s.
Since generally the presence of the eye-closure reflex should not be assumed, one
should close the eyes or immediately turn away if the laser radiation impinges on the
eye.
Lasers of Class 2M are not subject to notification and a laser protection officer is not
required. Mark the laser area recognizable and everlasting.
Safety
i
IMPORTANT!
The protection class is
limited to water (no
penetrating liquids or
similar).
i
IMPORTANT!
Follow laser protection
regulations.
WARNING!
Do not intentionally look
into the laser beam.
Consciously close the
eyes or turn away if the
laser radiation impinges
on the eye.
Содержание scanCONTROL2800