Page 22
Installation and Assembly
optoCONTROL CLS1000
4.5.4
Model CLS1000-AI
4.5.4.1 Pin Assignment
Pin Signal/assign-
ment
Description
Comments, circuitry
PC1000-x
cable color
1
2
3
4
1
2
3
4
5
1
V
+
Supply voltage
12 ... 30 VDC
I
max
< 50 mA
Brown
2
OUT2
Analog output
0 ... 20 mA / 4 ... 20 mA
switchable ≤
R
L
max 500 Ohm
White
Fig. 14 4-pin
cable socket,
view of solder
side
Fig. 15 5-pin cable
socket,
view of solder side
1
3
GND
Ground
Ground for supply, switch-
ing output Q and analog
output and trigger (IN)
1
Blue
4
OUT1
Switching out-
put Q1
V
+
max. 12 ... 30 VDC,
I
max
= 100 mA, Signal level
depends on supply volt-
age, switching behavior
programmable: NPN, PNP,
PP
Black
5
1
TRG
1
Trigger (IN)
1
max. 30 VDC; input current
I
max
= < 1 mA
Gray
1
1) Only for controller CLS1000-AI-xx-T with trigger function
Alternatively, for the model without trigger function, an angled signal/power supply cable PC1000/90-5 with a connector
angled at 90° (or open ends) is available with shielding.
4.5.4.2 Analog Output
The sensor supplies a current output of 0 ... 20 mA or 4 ... 20 mA.
Analog output properties
Sensor
pin
PC1000-x
cable color
Comments
Current output
0 ... 20 mA or 4 ... 20 mA;
2
White
Analog output
switchable ≤
R
L
max. 500 Ohm
3
Blue
Ground
Fig. 16 Analog output switching
4.5.4.3 Switching Output
The switching variants of the switching output (PP/PNP/NPN) as well as the switching thresholds can be changed and
adjusted using the operating menu.
The switching behavior (NPN, PNP, push-pull (PP)) of the switching output depends on the programming.
GND
1
3
4
Q
PP / PNP / NPN
0/4 mA
20 mA
2
U Analog
V
+
Brown
White
Black
Blue
Fig. 17 Circuit for CLS1000-AI