Page 25
optoCONTROL 1202
6.2
Evaluation Mode
The optoCONTROL is delivered with the mode L-EDGE factory-made. You can change the mode only via the
software tool.
L-EDGE:
The first detected edge in the
intensity profile of the CCD line is
evaluated.
SMR
EMR
SMR
EMR
R-EDGE:
The first detected edge in the
intensity profile of the CCD line is
evaluated.
SMR
EMR
SMR
EMR
WIDTH:
The difference between the last
edge and the first edge (for width
and gap) in the intensity profile of
the CCD is evaluated.
SMR
EMR
SMR
EMR
CENTER:
The mean value of the first and
second edge is used as measure-
ment value:
CENTER = (R-EDGE+L-EDGE)/2
SMR
EMR