Page 9
LEDs Controller
IMS5x00
LEDs Controller
Power on
Green
Operating voltage available
The LED’s Intensity and Range
flashes with their current color
during a synchronization error.
1) When measuring outside
the optimum current value of
the SLED, the controller will
measure, but the measure-
ment accuracy may not be as
specified.
Status
Off
No error
If EtherCAT is active, meaning of the LED is conform with the Ether-CAT guidelines.
Red
Signal in saturation
Yellow
Signal too low
Green
Signal ok
Red
SLED of
Yellow
SLED warms up
Green
SLED ready for operation
Yellow flashing
SLED current outside the optimal value range
1
Red
Pilot laser off
Green
Pilot laser on
Green
Pilot laser is alternately turned on and off,
if no target object or outside the measuring range
IMS5400-DS19
IMS5400-TH45
IMS5400-TH70
IMS5600-DS19
IMS5400-DS19/MP
IMS5400-TH45/MP
IMS5400-TH70/MP
IMS5600-DS19/MP
Distance and thickness measurment Distance measurment
Thickness measurment
Red
No target object, or target object
outside the measuring range
The expected number of peaks was not found or
it was not possible to assign a distance.
The expected number of peaks was
not found or it was not possible to
assign a thickness.
Yellow
Target close to mid of measuring
range
The expected number of peaks was found. A valid
distance could be found for each peak. Center
point of the target in the area surrounding the mid
of the measuring range.
-
Green
Target within measuring range
The expected number of peaks was found. A valid
distance could be found for each peak.
The expected number of peaks was
found. A valid thickness could be
found for each peak.