Page 17
Initial Operation
IMS5x00
Number of Peaks
Number of peaks of the FFT signal used for evaluation in distance and thickness measurement. The number of peaks
can be selected in the
Settings
>
Data recording
>
Number of peaks
.
This function is possible for the following systems:
- IMS5400-DS19/MP: max. 13 layers or 14 distance peaks
- IMS5400-TH45/MP: max. 5 layers
- IMS5400-TH70/MP: max. 5 layers
- IMS5600-DS19/MP: max. 13 layers or 14 distance peaks
Make sure to count the peaks in the correct manner, see Chap. Messpeak Sorting.
The standard systems IMS5400-DSxx, IMS5400-THxx and IMS5600-DSxx evaluate one layer.
i
If no distances need to be determined for a measurement, Micro-Epsilon recommends using an IMS5400MP-
THxx.
The material selection for a thickness measurement starts with the thinnest layer (layer 1) independent of the physical
arrangement in the measuring object. The material selection for a distance measurement corresponds to the actual
physical arrangement in the measuring object.