Page 8
Setup, Connection Options
IMS5x00
Pluggable Screw Terminals
Pin
Description
Comments
U/I out
GN
D
Shield
Analog Out
U/I out
Voltage output
0 ... 5 V; 0 ... 10 V;
R
i
appr. 50 Ohm
5.5 V / 10.9 V with error, outside measuring
range
Current output
4 ... 20 mA;
R
L
≤ 500 Ohm
23.7 mA with error, outside measuring range
GND
Ground analog output
Galvanically connected with supply
+Sync/Trig
-Sync/Trig
Synchronization input/out-
put, trigger input
RS422 level (EIA422)
+
Sync/T
rig
-
Sync/T
rig
Digital I/O
GND
Shield
TrigIn
GND
Er
ror 1
GND
Shield
Er
ror 2
GN
D
TrigIn
Trigger input
TTL or HTL level
TTL: Low ≤ 0.8 V, High ≥ 2 V
HTL: Low ≤ 3 V, High ≥ 8 V
Error 1 / 2
Switch outputs
NPN, PNP or Push-Pull
I
max
= 100 mA,
U
H max
= 30 V
GND
Ground potentials
All GND conductors are interconnected with
one another and to operating voltage ground.
24 VDC
Operating voltage
± 15 %,
I
max
< 1 A
24 VDC
GN
D
Shield
Power
GND
Operating voltage ground
GND is galvanically connected to GND of
switching outputs, synchronization, analog
and encoder input
Shield
Shields to respective output/input, connector housing
The plug-in screw terminals are designed for a conductor cross-section of 0.14 mm² up to 1.5 mm².