Page 17
Installation
optoNCDT 22xx
5.
Installation
The ILD 22xx is an optical sensor for measurements with micrometer accuracy. Make sure it is handled care-
fully when installing and operating. MICRO-EPSILON recommends the use of protective housings if the sen-
sor operates in a dirty environment or higher ambient temperature, see Chap.
14.2
.
CE1800-x
PC1800-3
PC1800-3/10/RS485 or
PC2200-x/USB/IND
PC2200-x/IF2008
System with sensor, sensor cable, controller and power supply and output cable
Fig. 4
Cable
Continuous high flex cable
Bending radius (min, permanent)
CE1800-x
•
50 mm
PC1800-x
•
60 mm
Optional Accessory:
CE1800-x Sensor cable extension
PC1800-x Power supply and output cable
x = cable length in m
Sensor cables with 5 m length or more must be included into the system calibration.