Page 53
Serial Interface RS422
optoNCDT 22xx
Stop Command
7.10
Name
STOP
Description:
Output of data via serial output is stopped
Format:
Response:
31
24 23
16 15
8 7
0
hex
31
24 23
16 15
8 7
0
hex
“+“
“+“
“+“
0x0D
0x2B2B2B0D
“I“
“L“
“D“
“1“
0x494C4431
“I“
“L“
“D“
“1“
0x494C4431
0xA0
0x76
0x00
0x02
0xA0760002
0x20
0x76
0x00
0x02
0x20760002
0x20
0x20
0x0D
0x0A
0x20200D0A
Get Sensor Settings
7.11
Name
Get_Settings
Description:
Reads out the sensor settings
Measuring frequency
-
0 = 10 kHz
1 = 5 kHz
2 = 2.5 kHz
3 = 20 kHz
Average number
-
is an integer
in hexadecimal code for mov-
ing and recursive average.
Note: The values for averag-
ing can be only a multiple of
(N = 2
n
).
For Median:
0 = 3
2 = 5
5 = 7
7 = 9
Flag
-
Hold Last Value
0 = hold last value not
1 = hold last value
Averaging method
-
0 = recursive
1 = moving
2 = Median
Offset value
-
Supplies the absolute value in
the moment zero setting was
done
Zero point
-
0 = absolute measurements
1 = relative measurements
Measuring range
-
is an
integer in hexadecimal
code in mm
Lock keys
-
0 = keys enabled
1 = keys locked
Digital data output
-
0 = data output off
1 = data output on
Laser state
-
0 = Laser off
1 = Laser on