Page 63
Set Sensor Parameter
optoNCDT 1320
Current output (with teaching)
Variables
Value range
Formula
I
OUT
= current [mA]
[3,8; <4] SMR reserve
[4; 20] measuring range
[>20; 20,2] EMR reserve
d
[mm]
=
16
*|n
[mm]
- m
[mm]
|
(I
OUT
[mA]
- 4)
MR = measuring range [mm]
{10/25/50/100}
m, n = teachig area [mm]
[0; MR]
d
= distance [mm]
[m; n]
Current output (with mastering and teaching)
Variables
Value range
Formula
I
OUT
= current [mA]
[3,8; <4] SMR reserve
[4; 20] measuring range
[>20; 20,2] EMR reserve d
[mm]
=
* |n
[mm]
- m
[mm]
|
(I
OUT
[mA]
- 12)
16
MR = measuring range [mm] {10/25/50/100}
MP = position object [mm]
[0; MR]
m, n = teaching area [mm]
for MP
£
0.5MR: [-MP; 0.5MR]
for MP > 0.5MR: [-0,5MR; MR - MP]
d
= distance [mm]
[m; n]