Page 57
Set Sensor Parameter
optoNCDT 1320
7.5.2.2
Characteristics Digital Output
Measurements are coded with 18 bit, if the based on zero setting or mastering. The master value itself can accept the double measu-
ring range. The examples below show the digital output behavior of an ILD1320-50 with 50 mm measuring range.
Target at 16 % measuring range
Target at 60 % measuring range
Target at 60 % measuring range
SMR
EMR
0 %
16 %
100 % MR
Target
10920
8.00 mm
SMR
EMR
0 %
60 %
100 % MR
Target
39183
30.00 mm
SMR
EMR
0 %
60 %
100 % MR
Target
39183
30.00 mm
Zero setting (master value = 0 mm)
Zero setting (master value = 0 mm)
Set master value 10 mm
SMR
EMR
0 %
16 %
100 % MR
Target
32760
0.00 mm
22483
-8.00 mm
86717
42.00 mm
EMR
0 %
60 %
100 % MR
Target
32760
0.00 mm
10 %
643
-30.0 mm
58454
20.00 mm
EMR
SMR
0 %
60 %
100 % MR
Target
45607
10.00 mm
7066
-20.0 mm
71301
30.00 mm
Digital minimum reached at 10 % MR