Page 67
Advanced Settings
confocalDT 24x1
6.13
Extras
Language/Sprache
Deutsch /
English
Language of the interactive web pages.
Factory
settings
Only reset
current setup
Check box Enables the user to replace the currently
used setup only.
Maintain interface
settings
Check box Allows to retain all Ethernet and RS422
interface settings without any changes.
Interface settings are hold when the controller is set to keep current Ethernet and RS422
interface settings.
6.14
Masking the Evaluation Range
Masking limits the range that the video signal uses for distance or thickness calculations.
This feature is used, for example, if ambient light with certain wavelengths (blue, red, IR)
causes video signal interference. It is also possible to mask the background if it reaches
into the measuring range.
Masking (start and end) is entered into the two boxes on the left (in %). The factory set-
tings are 0 % (start) and 100 % (end).
i
If you limit the video signal area, a peak is detected only, if it lies completely within
the masked area, i. e. above the threshold. The measuring range can be reduced
thereby.
Masked range
Fig. 56 Limiting the video signal