Page 140
Appendix |EtherCAT Documentation
confocalDT 24x1
Object 2250h: Exposure mode/Measuring rate
2250
RECORD Shutter mode/measuring rate
Subindices
0
VAR
Number of entries
4
Unsigned8
ro
1
VAR
Shutter mode
x
Unsigned8
rw
2
VAR
Measuring rate
x
Unsigned8
rw
3
VAR
Shutter time 1
x.xx
FLOAT32
rw
4
VAR
Shutter time 2
x.xx
FLOAT32
rw
5
VAR
Manual measuring rate x.xx
FLOAT32
rw
Further details can be found in the section Exposure mode / Measuring rate, see Chap.
and
Shutter mode:
0 - Automatic mode
1 - Measurement mode
2 - Manual mode
3 - Two-time mode alternating
4 - Two-time mode automatically
Measuring rate:
Setting value
IFC2451
IFC2461
IFC2471LED
IFC2471
0
-
70 kHz
70 kHz
1
-
50 kHz
50 kHz
2
-
25 kHz
25 kHz
25 kHz
3
10 kHz
10 kHz
10 kHz
10 kHz
4
5 kHz
5 kHz
5 kHz
5 kHz
5
2.5 kHz
2.5 kHz
2.5 kHz
2.5 kHz
6
1 kHz
1 kHz
1 kHz
1 kHz
7
0.3 kHz
0.3 kHz
-
0.3 kHz
8
0.2 kHz
0.2 kHz
-
-
9
0.1 kHz
0.1 kHz
0.1 kHz
-
10
Choice from subindice 5