Page 104
Appendix |ASCII Communication with Controller
confocalDT 24x1
A 5.4.3.5 Delete a Material
MATERIALDELETE <Name>
Delete a material
-
Name: Name of material (Length: max. 30 characters)
A 5.4.4 Number of Peaks, Material Settings Multilayer Measurements
A 5.4.4.1 Number of Peaks and Enabling/Disabling Refractive Correction
REFRACCORR on|off [<Number of peaks>]
On: The refractive index correction is performed with the adjusted materials, default set-
ting.
Off: The refractive index 1.0 is expected for all layers.
Number of peaks: Number of the used peaks expected for the multilayer measurement.
The number must be between 3 ... 6 (the distance and thickness measurement modes
are available for a few peaks).
A 5.4.4.2 Material Settings
MATERIALMP [<Material1>[<Material2>[<Material3>[<Material4>[<Mat
erial5>]]]]]
Displaying and setting the materials for the layers between the peaks 1 up to 6.
The existing material is maintained with input from „“.
MATERIALINFO [<Number of layer>]
Display the material for the opposite layer. The information is output to layer 1 without
parameter.
A 5.4.5 Measurement Value Processing
A 5.4.5.1 Averaging of Measurement Value
AVERAGE NONE|MOVING|RECURSIVE|MEDIAN [<Averaging depth>]
-
The averaging value always affects all to be output displacement and difference val-
ues.
-
NONE: No averaging value
-
MOVING: Moving averaging value (averaging depth 2, 4, 8, 16, 32, 64, 128, 256, 512
and 1024 possible)
-
RECURSIVE: Recursive averaging value (averaging depth 2 up to 32768 possible)
-
MEDIAN: Median (averaging depth 3, 5, 7 and 9 possible)
A 5.4.5.2 Spike Correction
SPIKECORR [ON|OFF[[<Number of evaluated measured values>][[<Tol-
erance range in mm>][<Number of corrected values>]]]
Spike correction is not enabled in the factory default settings.
Factory settings Min
Max
Number of measured values evaluated 3
1
10
Tolerance range in mm
0.1000000
0.0000000
100.0000000
Number of corrected values
1
1
100
The tolerance range is set in mm with seven decimals.
A 5.4.5.3 Selection of the Signal for the Statistics, Multilayer Measurement
STATISTICSIGNAL DIST1|DIST2|DIST3|DIST4|DIST5|DIST6|THICK12|
THICK13|THICK14|THICK15|THICK16|THICK23|THICK24|THICK25|THICK26|
THICK34|THICK35|THICK36|THICK45|THICK46|THICK56
Information of the signal for which the statistics calculation should be applied.
A 5.4.5.4 Setting the Statistics Calculation
STATISTICDEPTH ALL|2|4|8|...|8192|16384
Input on how many measurement values the statistics data minimum, maximum and
peak to peak are determined.