5.6 Mobile Oxygen Carrier (MOC)
•
Should you require maintenance of the regulator(s) used in the MOC, contact the manufacturer
of the regulator to arrange for service. Vendor information is provided with the MOC
regulator at the time of purchase.
•
Ensure the regulator(s) contain the Oxygen seal when replacing Oxygen cylinder(s).
•
Monitor pressure of the cylinder(s) regularly to ensure enough Oxygen supply is available.
•
For the dual carrier, it is recommended to have only one cylinder open when powering the
Life-Stat
®
. When the Oxygen level is low, the other cylinder can then be opened and the low
cylinder closed.
•
Refer to additional
Warnings
listed in the General Warnings, Cautions and Notes section on
the use of Oxygen-related equipment.
5.7 Periodic Preventive Maintenance
The following recommended preventive maintenance procedures and checks can help increase the life
of the Life-Stat
®
, its related accessories and assure they are in proper operating condition.
NOTE:
There are no user-serviceable parts inside the Life-Stat
®
and no calibrations
or adjustments are needed for routine use. However, the general readiness and
function of the system can, and should, be evaluated on a regular basis. These checks
are performed on three levels.
1.
Shift check
– A series of checks that should be done after each use and at the start of every
shift. (See procedure below.)
2.
Functional check
– A complete visual and functional check of the Life-Stat
®
. (See
procedure below.)
3.
Factory service
– A recalibration and routine maintenance of internal components performed
at the manufacturer by factory trained personnel after the device has delivered over 1,000,000
(1 million) compression cycles. This is indicated by the START/ON green LED flashing
continuously.
The schedule for performing these procedures should be determined by the user, taking into
consideration specific circumstances and frequency of use. Use the table below as a guide.