CAUTION: The Life-Stat
®
requires a medical grade Oxygen source capable of
delivering pressure from 50 to 90 psi (3.515 to 6.327 kgf/cm
2
), with a minimum flow
rate of at least 45 LPM. Always follow safe Oxygen handling practices with Oxygen
cylinders and regulators.
3.4.2 Life-Stat
®
Application to the Patient:
A.
Before inserting the Life-Stat
®
into the BackBoard, ensure the
Arm is raised to and locked near the top of the Column and
positioned so that the dome is towards the patient’s feet. NOTE:
When oriented in this position, the Life-Stat
®
is top-heavy and
prone to fall over. Grasp the Life-Stat
®
at the Base and Handle.
B. Insert the Base into the side slot of the BackBoard on whichever
side of the patient is most convenient.
C. With the Base fully inserted into the BackBoard and during a
pause in the manual CPR effort, loosen the Arm Lock and swing
the Arm over the patient’s chest locating the Massager Pad over
the sternum, as you would for the heel of your hand when
performing manual CPR.
D. Lower the Arm until the Massager Pad contacts the patient’s
chest. Then, apply slight downward pressure on the Arm to
position the Piston inside the Dome to align with the “-” mark on
the Dome. Tighten the Arm Lock Lever.
WARNING: The Massager Pad must not extend over the
xiphoid process. This could result in injury to the patient.