![Michigan Instruments LIFE-STAT 1008MII Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/michigan-instruments/life-stat-1008mii/life-stat-1008mii_instructions-for-use-manual_1784714022.webp)
REV: 2019-04
22
WARNING: Injury to patient may occur if Arm is adjusted too low,
as indicated by the top of the Piston moving up beyond “-” on the
Dome.
WARNING: Patient chest compressions may be insufficient to be
effective if Arm is adjusted too high as indicated by the top of the
Piston not moving up to “-” on the Dome.
WARNING: Patient is more likely to shift from optimum position
relative to Massager Pad if Arm is adjusted too high, as indicated by
the top of the Piston not moving up to “-” on the Dome.
CAUTION: If the Arm Lock Lever is not securely tightened, Arm
height or Massager Pad location may shift position relative to the
patient.
3.4.3 Determine Compression Depth:
Determine the depth of compression by referring to
the Sternal Deflection Number located on the scale
on the back of the Column. The white arrows
indicate the depth required to provide the 20% A-P
sternal deflection for the patient.
WARNING: Do NOT use the Life-Stat
®
if the white arrows indicate
in the red area of the scale.
NOTE:
AHA Guidelines recommend for adult patients, a minimum
compression depth of at least 2" (5cm). If the A-P chest diameter number
indicated on the Life-Stat
®
is less than 5, then deliver compressions
minimally to the depth indicated by the -5- marking on the dome.
3.4.4 Life-Stat
®
Activation:
WARNING: Failure to ensure that the COMPRESSION DEPTH knob
(Control #2) is turned fully counterclockwise upon initial application to the
patient and prior to selecting a CPR mode via the SYSTEM CONTROL #1 will
deliver compressions to the patient at the depth last set by Control #2. This
depth may not be set for the correct A-P Diameter and could possibly cause
serious damage or inhibit resuscitating the patient.
Содержание LIFE-STAT 1008MII
Страница 8: ...REV 2019 04 8 Blank Page Intentional...
Страница 28: ...REV 2019 04 28 Blank Page Intentional...
Страница 46: ...REV 2019 04 46 Blank Page Intentional...
Страница 48: ...REV 2019 04 48 Blank Page Intentional...