REV: 2019-04
25
3.4.6 Procedure to Interrupt (Suspend) Compressions:
To perform pulse checks or perform analysis with an AED (and/or defibrillate manually), simply press
the PAUSE button. This will interrupt compressions only, while ventilations continue at the previous
rate and volume setting. To resume compressions, press the desired mode button (30:2 or CCV). The
same depth of compression previously set by the COMPRESSION DEPTH knob will be delivered. If
stopping the delivery of ventilations is desired temporarily, rotate the VENTILATION VOLUME
knob fully counterclockwise. To resume ventilations, rotate the VENTILATION VOLUME knob
clockwise to the desired volume.
CAUTION: When the Life-Stat
®
is used in conjunction with automatic
external defibrillators (AEDs), or other therapeutic devices which must
utilize an ECG signal, interruption of the cardiac compressions as
described may be required to avoid the ECG motion artifact associated
with cardiac compressions.
The Life-Stat
®
is electrically insulated and should cause no interference during
routine cardiac monitoring or manual defibrillation. However, conductive fluids or
gels may provide stray current paths.
WARNING: DO NOT touch the Life-Stat
®
during defibrillation.
3.5 To Remove the Life-Stat
®
from the Patient:
A. Press the STOP/OFF mode button on the SYSTEM CONTROL (#1) keypad.
B. Turn COMPRESSION DEPTH (#2) knob fully counterclockwise.
C. Turn VENTILATION VOLUME (#3) knob fully counterclockwise.
D. Disconnect the O
2
Supply Hose first from the Life-Stat
®
O
2
Supply Connector by pressing the
release button. Then, disconnect the connector from the Oxygen source by pulling the collar
back from the connector to release it.
CAUTION: Disconnecting the Oxygen hose from the Oxygen source end
first may not allow the device to properly vent the internal pneumatic
ports. Always disconnect the O
2
Supply Hose from the Life-Stat
®
first
.
NOTE:
Upon detaching the O
2
Supply Hose from the Life-Stat
®
, an abrupt and
loud release of Oxygen from the Column Buffer will occur. This is intentional
and required to purge the Life-Stat
®
of its reserve Oxygen.
E. Remove the Breathing Hose/Non Re-breathing Valve from the patient and the PDV. Discard
the Breathing Hose. Discard the Non Re-breathing Valve if a single use type.