REV: 2019-04
23
WARNING: Failure to ensure that the VENTILATION VOLUME knob
(Control #3) is turned fully counterclockwise upon initial application to the
patient and prior to selecting a CPR mode via the SYSTEM CONTROL #1
will deliver ventilations to the patient (if the breathing hose is applied to the
patient) at the tidal volume last set by Control #3. This setting may not be
correct and could deliver either excessive or insufficient tidal volumes to the
patient.
With the SYSTEM CONTROL (#1) in the STOP/OFF position:
A.
Ensure:
a.
COMPRESSION DEPTH knob (#2) is rotated fully counterclockwise
b.
VENTILATION VOLUME knob (#3) is rotated fully counterclockwise
c.
The pressure Indicator at the top of the Column shows “green” indicating adequate O
pressure.
B.
Press the START/ON button on the SYSTEM CONTROL (#1) keypad. The system will
perform a quick self check. When the system is ready the green LED indicator next to the
START/ON button will illuminate.
C.
Choose the desired CPR mode by pressing the desired mode button (30:2 or CCV). The blue
LED indicator near the selected mode button will illuminate.
D.
Rotate COMPRESSION DEPTH knob (#2) clockwise until sufficient compression depth is
demonstrated by viewing the Piston at eye level. Increase the depth until the top of the Piston
reaches the A-P Diameter Sternal Deflection Number on the Dome corresponding to the Sternal
Deflection Number reading taken from the scale on the back of the Column. This will deliver the
recommended A-P Diameter for the patient.
WARNING: Injury to patient may occur if compression depth is set too
deep or the knob is inadvertently moved clockwise.
WARNING: Patient chest compressions may be insufficient if
compression depth is set too shallow or the knob is inadvertently
moved counterclockwise.
WARNING: With the Life-Stat
®
in use, care must be taken to prevent
kinking or collapsing of the O
2
Supply Hose.