Michigan Instruments LIFE-STAT 1008MII Скачать руководство пользователя страница 1

REV: 2019-04 

  

 

 

 

 

 

 

 

 
 
 

Model 1008

MII

  

Mechanical CPR System

 

Instructions for Use (IFU) Manual 

(Part Number 

h

 16005) 

 

 

     

 

     

 

 

 

 

 

 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Manufactured in the USA by:

 

 
 

 

 

4717 Talon Court SE 

     

 

 

 

          Grand Rapids, MI 49512 USA

 

            

 

 

    Tel:  (800) 530-9939 or (616) 554-9696     

 

 

 

 

      

 

 

        Fax:  (616) 554-3067 

        e-mail:  

[email protected]

 

        website:  

www.life-stat.com

 

Содержание LIFE-STAT 1008MII

Страница 1: ...System Instructions for Use IFU Manual Part Number h 16005 Manufactured in the USA by 4717 Talon Court SE Grand Rapids MI 49512 USA Tel 800 530 9939 or 616 554 9696 Fax 616 554 3067 e mail mii michiga...

Страница 2: ...e of manufacture appears under symbol in YEAR MO format on device Y Caution Attention Consult Accompanying Documents Operators are to refer to information provided with the device Direct Current 9V in...

Страница 3: ...or Appropriate Wall Access Adapter 15 2 7 The Carrying Storage Case 16 3 Setup and Operation 16 3 1 Precautions 17 3 2 Recommendations 18 3 3 Positioning the Patient 19 3 4 Life Stat Deployment 20 3 4...

Страница 4: ...38 5 12 Electromagnetic Environmental Declaration 39 5 13 Parts List Life Stat 42 6 Warranty and Factory Service Information 43 6 1 Model 1008MII Life Stat Cardiopulmonary Resuscitator Warranty Agreem...

Страница 5: ...h its operation Keep this manual in a location where it is available for quick reference The format is designed to allow each section to be scanned quickly for answers to specific questions The Table...

Страница 6: ...e Life Stat into your cardiac arrest management regimen of care CAUTION Current American Heart Association guidelines do not recommend the use of mechanical CPR on infants or children WARNING This dev...

Страница 7: ...he instructions for preventive maintenance and cleaning procedures after each use They are found in the Care Cleaning and Disinfection section of this manual CAUTION Submersion of the Life Stat in wat...

Страница 8: ...REV 2019 04 8 Blank Page Intentional...

Страница 9: ...ompressor is powered by compressed Oxygen while timing is controlled by a 9V electronic circuit The device is electrically insulated allowing it to be freely and safely used in conjunction with routin...

Страница 10: ...ompression is easily monitored using the markings on the Dome surrounding the Piston The Column also serves as a storage tank that holds sufficient Oxygen to continue to drive the Life Stat for severa...

Страница 11: ...ctivate the 30 compressions to 2 ventilations operating mode and the blue LED indicator next to the button will illuminate CCV button Pressing this button will activate the Continuous Compressions Ven...

Страница 12: ...the white arrows B are located Set the compression depth to the corresponding indicator number on the Dome C to match the A P diameter number indicated on the Column 3 VENTILATION VOLUME Control 3 Th...

Страница 13: ...properly seated 3 Re insert the tray and push in until an audible click is heard ensuring the tray is properly seated inside the compartment 4 Repeat for the battery located on the opposite side Impo...

Страница 14: ...ick is heard to ensure it is seated properly 5 Discard the used battery appropriately when convenient The opposing battery may be hot swapped also following the above steps at this time If it is not r...

Страница 15: ...compressed medical Oxygen It incorporates couplers on each end and a check valve to retain the Oxygen during a source change Wall adapters are available that connect to the O2 Supply Hose to allow co...

Страница 16: ...Cylinders not included 2 7 The Carrying Storage Case The Carrying Storage Case is constructed of a durable nylon The Life Stat O2 Supply Hose and code related supplies are stored in the Case in a mann...

Страница 17: ...rity with accepted medical practices in your area is very important 4 Personnel certified in manual CPR must always be present to monitor the patient during Life Stat operation in the event of a mecha...

Страница 18: ...eally positioned to expedite removal from the Carrying Storage case and application to the patient When placing the BackBoard under the patient ensure the cross strap is positioned under the patient s...

Страница 19: ...marking on the cross strap Ensure the cross strap is positioned under the patient s neck Connect the retaining straps over the shoulder and under the arm pit of the patient then buckle and tighten th...

Страница 20: ...le connector of the Oxygen source then release the collar to secure the connector Pull gently on the hose to ensure a secure connection F Attach the opposite end of the hose to the Life Stat O2 Supply...

Страница 21: ...nd prone to fall over Grasp the Life Stat at the Base and Handle B Insert the Base into the side slot of the BackBoard on whichever side of the patient is most convenient C With the Base fully inserte...

Страница 22: ...mber located on the scale on the back of the Column The white arrows indicate the depth required to provide the 20 A P sternal deflection for the patient WARNING Do NOT use the Life Stat if the white...

Страница 23: ...elf check When the system is ready the green LED indicator next to the START ON button will illuminate C Choose the desired CPR mode by pressing the desired mode button 30 2 or CCV The blue LED indica...

Страница 24: ...dvertently moved clockwise WARNING Patient ventilation may be insufficient if the ventilation volume is set too low or the knob inadvertently moved counterclockwise WARNING Should the blue rubber diap...

Страница 25: ...interference during routine cardiac monitoring or manual defibrillation However conductive fluids or gels may provide stray current paths WARNING DO NOT touch the Life Stat during defibrillation 3 5...

Страница 26: ...athing Valve h 14384 supplied with the device is intended for single use only Do not re use CAUTION It is very important to follow the instructions for preventive maintenance and cleaning procedures a...

Страница 27: ...ose on the Base for easy access 4 2 Shipping If the Life Stat must be shipped for any reason a factory carton with protective foam inserts is recommended to protect the device Replacement cartons are...

Страница 28: ...REV 2019 04 28 Blank Page Intentional...

Страница 29: ...external surfaces of the Life Stat O2 Supply Hose BackBoard Carrying Storage Case and related accessories to remove foreign matter after each use Discard single use items such as the Breathing Hose an...

Страница 30: ...y carefully examine the Hose Adaptor Lip Valve and Pressure Limit Alarm Discard any cracked or damaged parts and replace as necessary F Install the Pressure Limit Alarm on the PDV and tighten snugly P...

Страница 31: ...p increase the life of the Life Stat its related accessories and assure they are in proper operating condition NOTE There are no user serviceable parts inside the Life Stat and no calibrations or adju...

Страница 32: ...use 10 times per month After each use Semi annually Every 5 years Use based on a 20 minute average run time In addition to the procedures for the Shift Check and Functional Check checklists are also p...

Страница 33: ...the 6 position and tighten the Arm Lock Lever 5 Inspect the O2 Supply Hose for kinks cracks cuts worn hose or damaged connectors Connect the O2 Supply Hose to the Oxygen source first and then connect...

Страница 34: ...re Indicator shows green LOW BATT LED not blinking or illuminated 7 Ventilation is delivered at 800ml setting Pressure Limit Alarm activates 8 Control 3 rotates and is secure 9 O2 Supply Hose disconne...

Страница 35: ...n the Arm Lock Lever Attach the O2 Supply Hose to the Oxygen source first b Check the operation of the O2 Supply Hose by connecting and disconnecting the Life Stat end of the supply hose a few times T...

Страница 36: ...ard slots 4 Ventilation Compression test a Piston reads and Arm Lock Lever secures Arm in place b O2 Supply Hose connects and disconnects easily c Pressure Indicator shows green d Ventilation is deliv...

Страница 37: ...ntrol 3 No mode selected Select wither 30 2 or CCV mode No compressions with increase of Control 2 Control 2 knob not secured to valve shaft Verify Control 2 knob is secured to shaft No ventilation wi...

Страница 38: ...le 0 0 317 gal Deliver 95 5 Oxygen to the patient Inspiratory Time 1 0 0 25 seconds Inspiratory Flowrate 0 to 50 0 2 5 LPM 0 to 13 20 0 66 gal Inspiratory Expiratory I E ratio 30 2 1 1 5 100 compressi...

Страница 39: ...tat is intended for use in the electromagnetic environment specified below The customer or the user of the Life Stat should assure that it is used in such an environment Immunity test IEC 60601 test l...

Страница 40: ...es NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a The ISM industrial scientific and m...

Страница 41: ...able Not Applicable 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be determined using the equation applicable to the fre...

Страница 42: ...ble 5 pkg 14682 Pressure Limit Alarm 14781 PDV Replacement Lip Valve 14785 PDV Replacement Cap Hose Adaptor Oxygen Management 15290 MOC Dual D Tank Soft Case 15300 MOC Dual E Tank Soft Case 15040 Soft...

Страница 43: ...OR CONSEQUENTIAL DAMAGES or any other expense liability beyond the furnishing of materials and labor for the repairs covered hereby This warranty does not cover mars and blemishes scratches or dents w...

Страница 44: ...related questions should also be directed to the Service Department C If your Life Stat CPR System must be returned to Michigan Instruments please observe the following procedures 1 First and foremos...

Страница 45: ...Stat CPR System that requires authorized factory service Michigan Instruments reserves the right to install used refurbished components that meet or exceed original manufacture specifications when per...

Страница 46: ...REV 2019 04 46 Blank Page Intentional...