XTC601 User’s Manual
Michell Instruments
37
INSTALLATION
4.4
Mechanical Installation
WARNING: Unused cable entries must be appropriately
blanked to maintain the integrity of the Exd casing.
There are 3 cable entries on the bottom surface, towards the rear, that can be utilized
by the customer in a variety of manners. The following standard options are available:
Cable Gland, Conduit Entry, Blanking Plug or Light Guide.
Figure 36
XTC601 Gas Connections and Cable Entries
4.4.1 Gas Connection Tape
The gas connections are on the bottom surface towards the front of the unit, these are
1/4" NPT for GP1 models and 1/8" NPT for Ex and GP2 models.
For applications where the oxygen concentration is
classed as enriched (>21% O
2
) any PTFE tape used must
be unsintered. This is to prevent an explosion due to
conventional PTFE tape acting as a potential fuel source.
Unsintered PTFE tape is available as an accessory from Michell Instruments (PTFE-
TAPE-02).
Содержание XTC601
Страница 1: ...XTC601 Binary Gas Analyzer User s Manual 97400 Issue 6 1 November 2020...
Страница 52: ...XTC601 User s Manual 42 97400 Issue 6 1 November 2020 APPENDIX A Appendix A Technical Specifications...
Страница 56: ...XTC601 User s Manual 46 97400 Issue 6 1 November 2020 APPENDIX B Appendix B Dimensional Drawings...
Страница 58: ...XTC601 User s Manual 48 97400 Issue 6 1 November 2020 APPENDIX C Appendix C Thermal Conductivity Table...
Страница 66: ...XTC601 User s Manual 56 97400 Issue 6 1 November 2020 APPENDIX E Appendix E Modbus Register Map HCG version...
Страница 73: ...XTC601 User s Manual Michell Instruments 63 APPENDIX F Appendix F Hazardous Area Certification...
Страница 80: ...http www michell com A PST Brand www ProcessSensing com...