XTC601 User’s Manual
54
97400 Issue 6.1, November 2020
APPENDIX D
Register Type I – Status/Error
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
r
r
r
r
r
r
r
r
r
r
r
r
r
r
r
r
Bit
HEX
Meaning
1
0002
%Gas out of range (beyond calibration range, e.g. 0–25%)
2
0004
Low alarm ON
3
0008
High alarm ON
4
0010
Ext Comp i/p signal error
(input < 3.6mA or > 21mA)
5
0020
Ext sens. signal error
(input < 3.6mA or > 21mA)
6
0040
Cell T not stable
(not within ± 0.15°C of setpoint for continuous period of 15 minutes)
7
0080
Cell T sensor error
(cell tempr measures <-50 or >80°C)
8
0100
NA
9
0200
Thermal Conductivity sensor error
(Vcomp <=1 or >=8191)
10
0400
PCB tempr too high
(PCB tempr > Cell tempr setpoint)
Register Type J
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w
eg Day
eg Month
For reading each 8 bits represents a RTC value. For setting only the Lower 8 bits are used for each
RTC value.
Register Type K: -32767 to +32767
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w
Range
= 0 to 65535 represents values
Conversion: (RegValue – 32767)
Содержание XTC601
Страница 1: ...XTC601 Binary Gas Analyzer User s Manual 97400 Issue 6 1 November 2020...
Страница 52: ...XTC601 User s Manual 42 97400 Issue 6 1 November 2020 APPENDIX A Appendix A Technical Specifications...
Страница 56: ...XTC601 User s Manual 46 97400 Issue 6 1 November 2020 APPENDIX B Appendix B Dimensional Drawings...
Страница 58: ...XTC601 User s Manual 48 97400 Issue 6 1 November 2020 APPENDIX C Appendix C Thermal Conductivity Table...
Страница 66: ...XTC601 User s Manual 56 97400 Issue 6 1 November 2020 APPENDIX E Appendix E Modbus Register Map HCG version...
Страница 73: ...XTC601 User s Manual Michell Instruments 63 APPENDIX F Appendix F Hazardous Area Certification...
Страница 80: ...http www michell com A PST Brand www ProcessSensing com...