XTC601 User’s Manual
62
97400 Issue 6.1, November 2020
APPENDIX E
Register Type K: -32767 to +32767
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w
Range = 0 to 65535 represents values
Conversion: (RegValue – 32767)
Register Type L: Alarm Configuration
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w
Bits
HEX
Meaning
0, 1
0003
Phase 1, Alarm 1 type – 00=OFF, 01=LOW, 10=HIGH
2, 3
000C
Phase 2, Alarm 1 type – 00=OFF, 01=LOW, 10=HIGH
4, 5
0030
Phase 3, Alarm 1 type – 00=OFF, 01=LOW, 10=HIGH
6, 7
00C0
Phase 1, Alarm 2 type – 00=OFF, 01=LOW, 10=HIGH
8, 9
0300
Phase 2, Alarm 2 type – 00=OFF, 01=LOW, 10=HIGH
10, 11
0C00
Phase 3, Alarm 2 type – 00=OFF, 01=LOW, 10=HIGH
Содержание XTC601
Страница 1: ...XTC601 Binary Gas Analyzer User s Manual 97400 Issue 6 1 November 2020...
Страница 52: ...XTC601 User s Manual 42 97400 Issue 6 1 November 2020 APPENDIX A Appendix A Technical Specifications...
Страница 56: ...XTC601 User s Manual 46 97400 Issue 6 1 November 2020 APPENDIX B Appendix B Dimensional Drawings...
Страница 58: ...XTC601 User s Manual 48 97400 Issue 6 1 November 2020 APPENDIX C Appendix C Thermal Conductivity Table...
Страница 66: ...XTC601 User s Manual 56 97400 Issue 6 1 November 2020 APPENDIX E Appendix E Modbus Register Map HCG version...
Страница 73: ...XTC601 User s Manual Michell Instruments 63 APPENDIX F Appendix F Hazardous Area Certification...
Страница 80: ...http www michell com A PST Brand www ProcessSensing com...