2.13.1 Transmitter Mounting - Sample Block (Optional)
The following procedure must be carried out by a
quali
fi
ed installation engineer.
To mount the transmitter into the sensor block (preferred method), proceed as follows,
refer to
Figure 18.
1.
Ensure that the protective cover (2), and its desiccant capsule (2a), have
been removed from the tip of the transmitter.
2.
Fit the bonded seal (4) over the threaded part of the transmitter body.
WARNING: Under no circumstances should the sensor
guard be handled with the
fi
ngers.
3. Screw the transmitter (1) into the sample block (3) and tighten to a
minimum torque setting of 30.5 Nm (22.5 ft-lbs).
NOTE: Use the
fl
ats
of the hexagonal nut and not the sensor body.
4.
Fit the transmitter cable/connector assembly to the plug located on the
base of the transmitter and tighten the
fi
xing screw (see Section 2.3).
1
2
3
2a
4
Figure 18
Transmitter Mounting - Sensor Block