Easidew Online User’s Manual
34
97094 Issue 18.5, March 2016
OPERATION
3.4.3 Alarm Set-Point Limit Con
fi
guration
P
SET
SET
x4
SET
SET
P
Main display
Main display
Set alarm setpoint lower limit
Set span alarm setpoint upper limit
The following procedure is used to set limits to
which the alarm levels can be set (usually after re-
con
fi
guring the instrument’s range for Fahrenheit
readings).
Figure 28 shows the operational key sequence.
1. Press
the
P
key once, the display will read
tECH
.
2. Press
the
SET
key once and the display will
fl
ash between
ConF
and
PinP
.
3. Press
the
key four times and the display
will
fl
ash between
ConF
and
GEnn
.
4. Press
the
SET
key once, the display will read
SU-L
.
5. Press
the
key once to display a
fl
ashing
4 digit number representing the current
minimum alarm level setting. (The default
setting for the °C range is -100.0).
6. Use
the
and
keys to set the required
new value (e.g. -148.0).
7. Press
the
SET
key to accept the new value.
The display will read
SU-u
.
8. Press
the
key once to display a
fl
ashing
4 digit number representing the current
maximum alarm level setting. (The default
setting for the °C range is 020.0)
9. Use
the
and
keys to set the required
new value (e.g. 068.0).
10. Press
the
SET
key to accept the new value,
followed by the
P
key to return to the main
display.
Figure 28
Set-up Alarm Set-Point Limits
Содержание Easidew Online
Страница 52: ...Easidew Online User s Manual 44 97094 Issue 18 5 March 2016 APPENDIX A Appendix A Technical Specifications ...
Страница 58: ...Easidew Online User s Manual 50 97094 Issue 18 5 March 2016 APPENDIX C Appendix C EC Declaration of Conformity ...
Страница 59: ...Easidew Online User s Manual Michell Instruments 51 APPENDIX C Appendix C EU Declaration of Conformity ...
Страница 67: ...Easidew Online User s Manual Michell Instruments 59 NOTES ...
Страница 68: ...http www michell com ...