INSTALLATION
2.11
Cable Connection
When installing the connector, and to ensure that full ingress protection is achieved, the
securing screw (with the O-ring and washer) must be tightened to a minimum torque
setting of 3.4 Nm (2.5 ft-lbs). The transmitter cable used must be a minimum diameter
of 4.6mm (0.2”).
1
2
3
4
4
Figure 15
Connector Installation
2.12 Electrical
Schematic
NOTE: The screen/shield should be connected for maximum performance
and to avoid interference.
3
1
Max Load
250R @ 12V
500R @ 24V
Supply
12V Min
28V Max
+
Screen/
Shield
Figure 16
2-Wire Connection Diagram