INSTALLATION
2.14 Signal Output Connections
The Easidew Online system has three signal outputs, Alarm 1 (ALr1), Alarm 2 (ALr2)
and the re-transmitted input signal (4-20 mA or 0-20 mA current loop signal depending
upon instrument con
fi
guration).
Figure 22 shows the relevant rear panel connections. Table 3 shows a summary of all
the electrical connections to the monitor.
The signal outputs will be connected to external systems that
can potentially in
fl
uence the operation of the process.
Alarm level signals could be at mains potential so it is essential
that, before connecting these signal lines, checks are made
to ensure that these inputs are not live and that it is safe to
handle them.
Alarm Outputs
Alarm 1 is a single pole make contact. Connect incoming signal lines to terminal 16
(common) and terminal 17 (normally open).
Alarm 2 comprises a set of changeover contacts. Connect incoming signal lines to
terminal 9 (common), terminal 8 (normally open) and terminal 7 (normally closed).
Figure 22
Monitor Rear Panel Connections
Содержание Easidew Online
Страница 52: ...Easidew Online User s Manual 44 97094 Issue 18 5 March 2016 APPENDIX A Appendix A Technical Specifications ...
Страница 58: ...Easidew Online User s Manual 50 97094 Issue 18 5 March 2016 APPENDIX C Appendix C EC Declaration of Conformity ...
Страница 59: ...Easidew Online User s Manual Michell Instruments 51 APPENDIX C Appendix C EU Declaration of Conformity ...
Страница 67: ...Easidew Online User s Manual Michell Instruments 59 NOTES ...
Страница 68: ...http www michell com ...