4. Infusion Pump Testing
This chapter details procedures for conducting each of the QA-IDS
tests.
4.1
Introduction
Infusion pump testers calculate flow rates from a measured time pe-
riod to fill a defined volume. If it takes one hour to fill a defined
volume of 10 milliliters, the flow rate is calculated to be 10 ml/hr.
This flow rate is called the
instant flow rate.
The
minimum volume
detection
of this measurement is 10 milliliters, since it needs that
amount to calculate a flow rate. The next instant flow rate calcula-
tion may be done when an additional volume of 10 ml is delivered
to the measuring device. When two or more instant flow rates have
been obtained, the calculation of a
mean flow rate
can be done by
averaging the instant flow rates.
Detecting unexpected flow changes requires detection of tiny infu-
sion volumes, as flow measurements may only be calculated every
time the defined volume is filled up. The ability to detect small infu-
sion volume requires that measurements be made frequently, hence
a high flow-sampling rate. The ability to detect only larger infusion
volumes decreases the sampling rate, as well as the ability to accu-
rately detect flow changes. International standards prescribe flow
sampling to be done every thirty seconds for all flow rates to ensure
a steady rate of infusion from all types of infusion devices. To fulfill
this requirement, minimum volume detection of the measuring de-
vice is critical.
Example:
To test an infusion pump for unexpected flow changes at a flow rate
of 10 ml/hr, with an accuracy of ± 2%, the following minimum flow
detection is needed:
Expected flow:
10 ml/hr = 166.7 pl pr. minute (divide by 60) = 83.3 µl
pr. 30 sec. (divide by 2).
2% of expected flow:
83.3 µl pr. 30 sec. x 2% = 1.67 µl pr. 30 sec.
A volume detection of 1.67 µl is required to detect a 2% unex-
pected, sudden flow change from an expected flow rate of 10 ml/hr
within 30 seconds.
The minimum volume detection of QA-IDS is 0.72 µl. That means
that less than 1 µl is needed for QA-IDS to detect any sudden flow
change between 0.10 ml/hr and 1000.0 ml/hr. QA-IDS performs
flow-samplings every second for all flow rates.
4.2 Test Preparation
23
Содержание QA-IDS
Страница 1: ...1 Q A ID S U s e r S e r v i c e M a n u a l Q A I D S I n f u s io n P u m p T e s t e r...
Страница 6: ...This page intentionally left blank 6...
Страница 12: ...This page intentionally left blank 12...
Страница 32: ...This page intentionally left blank 32...
Страница 38: ...This page intentionally left blank 38...
Страница 40: ...This page intentionally left blank 40...
Страница 41: ...Processor board Component Location 41...
Страница 42: ...6 10 1 Processor board Schematic Diagram 1 42...
Страница 43: ...6 10 2 Processor board Schematic Diagram 2 Digital Part 43...
Страница 44: ...6 10 3 Processor board Schematic Diagram 3 CPU 44...
Страница 45: ...6 10 4 Processor board Schematic Diagram 4 Memory 45...
Страница 46: ...6 10 5 Processor board Schematic Diagram 5 RS232C and Printer 46...
Страница 47: ...6 10 6 Processor board Schematic Diagram 6 Digital Outputs 47...
Страница 48: ...6 10 7 Processor board Schematic Diagram 7 Digital Inputs 48...
Страница 49: ...6 10 8 Processor board Schematic Diagram 8 A D 49...
Страница 50: ...6 10 9 Processor board Schematic Diagram 9 Stepper Motor and Drivers 50...
Страница 51: ...6 10 10 Processor board Schematic Diagram 10 Pressure Sensor and Temper ature 51...
Страница 52: ...52...
Страница 53: ...6 10 11 Processor board Schematic Diagram 11 Main Part 53...
Страница 54: ...54...
Страница 55: ...6 10 12 Processor board Schematic Diagram 12 Keyboard Pressure Sensor and Pump Index 55...
Страница 56: ...56...
Страница 57: ...6 10 13 This page intentionally left blank 57...
Страница 59: ...This page intentionally left blank 59...
Страница 61: ...This page intentionally left blank 61...
Страница 62: ...62...