46
S
CALE
O
PERATION
C
ALIBRATION
Scale Calibration Procedure (lbs. or kg) with Remote Display
7. Make sure there is no load on the scale platter.
8.
Calibrate a zero load.
Press the right arrow (
►
) button once when the remote display reads CAL 1 and there is
no
load
on the scale platter.
Figure 44.
9. Place a half load (15.00 lbs. or 7.500 kg) on the center of the scale platter then press the right arrow (
►
) button.
Figure 45.
10. Add an additional half load (15.00 lbs. or 7.500 kg) to the existing half to simulate a full load, center entire load then press
the right arrow (
►
) button.
Figure 46.
11. Remove half of the load (15.00 lbs. or 7.500 kg), center the remaining load then press the right arrow (
►
) button.
Figure 47.
Содержание StratosS MS2220
Страница 1: ...METROLOGIC INSTRUMENTS INC MS2220 StratosS Scanner Avery Scale Installation and User s Guide...
Страница 6: ...v...
Страница 14: ...8 BASE MODEL CHARACTERISTICS MS2220 Scanner Avery Scale Components Figure 7 MS2220 Components...
Страница 86: ......
Страница 87: ......
Страница 88: ...December 2006 Printed in the USA 0 0 0 2 0 3 5 C...