Problem
Cause
Remedy
Individual peaks are
greater than expec-
ted.
Sample – Sample carry-
over from previous meas-
urements.
Check the rinsing time
The background
conductivity is too
high.
The incorrect eluent is
being used.
The retention times
are poorly reprodu-
cible.
The eluent path has a leak.
Check all of the connections along the eluent
path and fix the leak.
The eluent path is blocked.
Check the eluent path and eliminate the block-
age.
The eluent contains gas
bubbles.
■
Check the connectors of the eluent degas-
ser .
■
Deaerate the high-pressure pump
Chromatograms
have poor resolution
Separation column –
Diminished separating effi-
ciency.
■
Regenerate the separation column
.
■
Replace the separation column
necting the separation column", page 41)
.
Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)
Capillary connections –
Dead volume in the sys-
tem.
Check the capillary connections
(use PEEK capillaries with an
inner diameter of 0.25 mm between the injec-
tion valve and detector).
Guard column – Dimin-
ished efficiency.
Replace the guard column
Separation column – Dead
volume at the column
head.
■
Install the separation column in the oppo-
site flow direction (if the leaflet says this is
permissible) and rinse into a beaker.
■
Replace the separation column
necting the separation column", page 41)
.
Precision problems -
the measured values
are highly scattered.
Injection valve – Sample
loop.
Check the installation of the sample loop
"Optional: Exchanging the sample loop",
page 29)
Sample – There are gas
bubbles in the sample.
Use the sample degasser .
Sample – The rinsing vol-
ume is too small.