![Metrohm 940 Professional IC Vario TWO Скачать руководство пользователя страница 66](http://html1.mh-extra.com/html/metrohm/940-professional-ic-vario-two/940-professional-ic-vario-two_manual_1779626066.webp)
5.8 Servicing the high-pressure pump
■■■■■■■■■■■■■■■■■■■■■■
58
■■■■■■■■
940 Professional IC Vario TWO (2.940.2100)
1
Release the coupling from the pressure screw and seal it with a stop-
per.
2
Unscrew and remove the pressure screw on the pump head's outlet
3
Loosen and remove the four fastening screws
key.
4
.
Servicing the piston
Carry out the following work on both pistons in turn.
Servicing a piston consists of the following tasks:
1.
Replace the piston seal.
2.
Clean or replace the zirconium oxide piston.
3.
Reinstall the piston.