3.5 Connecting the drainage tubing and leak sensor
■■■■■■■■■■■■■■■■■■■■■■
20
■■■■■■■■
940 Professional IC Vario TWO (2.940.2100)
Connecting the drainage tubing
1
Cut a piece of silicone tubing into three pieces using scissors: 2 ×
approx. 40 cm and 1 × 20 cm.
2
Attach one end of the 40 cm long piece to the drainage tubing con-
nection on the bottle holder.
3
Attach one end of the 20 cm long piece to the drainage tubing con-
nection on the detector chamber.
4
Attach each of the loose ends of both pieces of silicone tubing to
one end of the Y connector.
5
Attach one end of the second 40 cm long piece to the third end of
the Y connector.
Attach the loose end to the right-side drainage tubing connection on
the base tray.
6
Attach one end of the second piece of silicone tubing to the left-side
drainage tubing connection on the base tray.