3 Installation
940 Professional IC Vario TWO (2.940.2100)
■■■■■■■■
41
Connecting the separation column
1 Removing the stoppers
■
Remove the stoppers from the separation column.
2 Installing the inlet of the separation column
CAUTION
When inserting the column, ensure that it is inserted correctly
based on the marked flow direction.
There are three options:
■
Attach the column inlet directly onto the guard column or,
■
if the guard column is connected to the separation column using
a connection capillary: Connect the column inlet to the guard col-
umn outlet capillary using a PEEK pressure screw (6.2744.070) or,
■
if no guard column is used (not recommended): Connect the col-
umn inlet capillary to the inlet of the separation column using a
short pressure screw (6.2744.070).
3 Rinsing the separation column
■
Place a beaker under the outlet of the separation column.
■
Start manual control in MagIC Net and select the high-pressure
pump:
Manual
▶
Manual control
▶
Pump
–
Flow
: Increase gradually up to the flow rate recommended
in the column leaflet.
–
On
■
Rinse the separation column with eluent for approx. 10 minutes.
■
Stop the high-pressure pump in the manual control in MagIC Net
again:
Off
.
4 Removing the coupling
■
Remove the coupling (6.2744.040) from the column outlet capil-
lary.
5 Installing the outlet of the separation column
■
Fasten the column outlet capillary to the column outlet using a
short PEEK pressure screw (6.2744.070).