3 Installation
940 Professional IC Vario TWO (2.940.2100)
■■■■■■■■
35
3.18
Initial start-up
Even before the guard column and separation column are installed, the
entire system must be completely rinsed with eluent for the first time.
Rinsing the IC system
CAUTION
The separation column and the guard column are not permitted to be
installed at the time of the initial start-up.
Make sure that a coupling (6.2744.040) is being used instead of the
columns.
1 Preparing the software
■
Start the
MagIC Net
computer program.
■
Open the
Equilibration
tab in MagIC Net:
Work-
place
▶
Run
▶
Equilibration
.
■
Import (or create) a suitable method.
Also see:
MagIC Net Tutorial
and online help.
2 Preparing the instrument
■
Ensure that the eluent aspiration tubing is immersed in the eluent
and that there is enough eluent in the eluent bottle.
■
Switch on the instrument.
MagIC Net detects the instrument and all of its modules.
3 Starting equilibration
■
Start the equilibration in MagIC Net:
Workplace
▶
Run
▶
Equili-
bration
▶
Start HW
.