1.2 Development of a method
736 GP Titrino: Getting started
6
Press <print> <reports> <enter> .
Your printout will look as follows:
Identification of the report type
(cu = curve)
Start volume and method identification
Scaling of curve axis
- Identification of the report type
- Initial pH and method with identification
- Volume and pH value of EP1
- Calculated result with result text
Entry of a stop criterion
Entry of a stop criterion
You don't want to stop your titration manually each time. Let
us enter an automatic stop criterion.
2 x <parameters>
<enter>
<enter>
<enter>
11.5 <enter>
2 x <QUIT>
Press <parameters> twice to display the title ">stop condi-
tions".
The absolute stop volume. It can be used as a safety volume
to prevent overflow of the titration vessel.
Enter a value for the stop volume that appears suitable for
your titration vessel.
Set the pH value, e.g. 11.5 as stop criterion.
If several stop criteria have been set, the one that is reached
first, applies.
Confirm the input with <enter> and quit the inquiry with
<QUIT>.
Try another titration with this method.
'cu
736 GP Titrino 05325 736.0010
date 1997-12-16 time 09:07 1
start V 0.000 ml DET pH ********
1.0 ml/div dpH=2.0/div
−−−−−−−−−−−−−−−−
'fr
736 GP Titrino 05325 736.0010
date 1997-12-16 time 09:07 1
pH(init) 2.88 DET pH ********
smpl size 2 ml
EP1 2.083 ml 7.64
RS1 3.80 g/l
manual stop
−−−−−−−−−−−−−−−−
>stop conditions
stop pH OFF
stop V 99.99 ml
stop V: abs.