2.2
Mode DET and MET, parameters
736 GP Titrino: Titration, dosing and measuring modes
14
2.2 Mode DET and MET, parameters
Display
Meaning
Input range
>titration parameters
meas.pt.density
min.incr.
[V step]
titr.rate
signal drift
equilibr.time
start V:
start V
factor
dos.rate
pause
dos.element:
meas.input:
temperature
Measuring point density.
Minimum increment for DET.
Volume increment for MET.
Dosing rate for increments.
Drift for measured value acquisition.
Waiting time for value acquisition.
Type of start volume: absolute, relative.
Volume for
absolute
start volume.
Factor for
calculation of
relative
start
volume: factor * smpl size.
Dosing rate for start volume.
Waiting time, e.g. after start volume.
Selection of the dosing unit.
Measuring input for pH and U.
Titration temperature.
0...
4
...9
0...
10.0
...999.9 ul
0...
0.10
...9.999 ml
0.01...150 ml/min,
max.
pH, U:0.5..
50
..999 mV/min, OFF
0...
26
...9999 s, OFF
abs., rel.,
OFF
0
...999.99 ml
0
...±999 999
0.01...150 ml/min,
max.
0
...999 999 s
internal D0,
external D1/D2
1
, 2, diff.
-170.0...
25.0
...500.0 °C
>stop conditions
stop V:
stop V
factor
stop pH
stop EP
filling rate
Type of stop volume: absolute or relative.
Volume for
absolute
stop volume.
Factor for
calculation of
relative
stop
volume: factor * smpl size.
Stop at measured value pH, U, I.
Stop after a number of EP's have been
found.
Filling rate after the titration.
abs.
, rel.,
OFF
0...
99.99
...9999.99
ml
0...
±999 999
pH: 0.00...±20.00,
OFF
U: 0...±2000 mV,
OFF
1...
9
, OFF
0.01...150 ml/min,
max.
>statistics
status:
mean n=
res.tab:
delete n=
Status of statistics calculation.
Number n of single values for statistics.
Result table for statistics calculation.
Delete data from sample number n.
ON,
OFF
2
...20
original
, delete n, delete all
1
...20
>evaluation
EPC
EP recognition:
fix EP1 at pH
pK/HNP:
Endpoint criterion.
Type of EP recognition: all EP's, only
greatest EP, only last EP, only EP's in
windows, no evaluation. If ”window” is
selected, lower and upper limits of
windows are inquired.
Interpolation of volume at given pH, U, or I,
resp. Up to 9 fix EP's.
Evaluation of pK or half neutralization
potential (HNP).
DET
0...
5
...200
MET
pH: 0.10...
0.50
...9.99
U: 1...
30
...999 mV
all
, greatest, last, window, OFF
pH: 0.00...±20.00,
OFF
U: 0...±2000 mV,
OFF
ON,
OFF
>preselections
Preselections for the sequence see DOS,
page 19