![Metrohm 736 GP Titrino Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/metrohm/736-gp-titrino/736-gp-titrino_instructions-for-use-manual_1779554007.webp)
1.2 Development of a method
736 GP Titrino: Getting started
5
Entry of the sample size
Entry of the sample size
4 x <smpl data>
2 <enter>
<select>
<enter>
Press <smpl data> repeatedly until "smpl size" appears
in the display.
Enter 2 (= 2 ml).
Use <select> to select the unit "ml" and confirm the
new value with <enter>.
The result is now calculated and can be displayed in place
of the equivalence point. If your method already includes a
formula at the end of the titration, the calculated result is
displayed directly after the titration. As we have entered the
formula later, we now have to select the result display:
Display of the result
Display of the result
<select>
<enter>
Press <select> repeatedly until ">display results" appears
in the display.
Press <enter> to move on to the result display.
Selection of the automatic report output
Selection of the automatic report output
If you have a printer connected, you can have the curve
and a result report printed out automatically at the end
of the titration.
4 x <def>
<enter>
<select>
<enter>
<QUIT>
Press <def>repeatedly until the display shows ">report".
Press <enter> to move to the definition of reports.
With <select> you select the individual report blocks
Use a ";" as delimiter between the report blocks. If you
wish to print out a curve and a full result report, enter
"curve;full".
Confirm the entry with <enter> and quit the inquiries
with <QUIT>
>display results
>report
smpl unit: ml
smpl size 1.0 g
%
RS1 3.80 g/l
report:curve;full
report: