![Metrohm 736 GP Titrino Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/metrohm/736-gp-titrino/736-gp-titrino_instructions-for-use-manual_1779554020.webp)
2.5
Mode STAT, parameters
736 GP Titrino: Titration, dosing and measuring modes
18
Display
Meaning
Input range
>monitoring
meas.val:
low lim.pH
up lim.pH
action:
rate:
temperature:
assign output:
L4 output:
Monitoring of measured values.
Limits for measured values.
Action, if a limit is exceeded. End: Abort.
Wait: Dosing interrupted, until measured
value is within limits again, then continue.
Monitoring of rate (same actions as
above).
Monitoring of temperature (same actions
as above).
Assignment of output L4 for monitoring.
Set a signal when values are out of limit.
ON,
OFF
pH: 0.00...±20.00
(0...±2000 mV, 0.0...±200.0 uA)
end, hold, wait,
none
ON,
OFF
ON,
OFF
meas, temp, rate, all,
none
active,
pulse
>preselections
req.ident:
req.smpl size:
display rate:
activate pulse:
Request of identifications after start.
Request of sample size after start.
Display of rate.
Pulse output on I/O line L6.
id1, id1 & 2, all,
OFF
Value, unit, all,
OFF
ON,
OFF
ON,
OFF