![Metrohm 736 GP Titrino Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/metrohm/736-gp-titrino/736-gp-titrino_instructions-for-use-manual_1779554030.webp)
3.3
Configuration, key <configuration>
736 GP Titrino: Other functions
28
3.3 Configuration, key <configuration>
Display
Meaning
Input range
>peripheral units
send to:
balance:
record:
Settings of peripheral units.
Selection of printer.
Selection of balance.
Selection of record at the analog output.
Epson, Seiko, Citizen, HP,
IBM
Sartorius
, Mettler, Mettler AT, AND,
Precisa
U
, dU/dt, V, dV/dt, U(rel), T
>auxiliaries
dialog:
date
time
run number
auto start
start delay
dev.label
program
General settings.
Selection of dialog language.
Current run number for result output.
Automatic starts of titrations.
Waiting time before start of titration.
Device label.
Program version.
english
, deutsch, français, español,
portuguese, italiano, svenska
YYYY-MM-DD
HH:MM
0
...9999
1...9999,
OFF
0
...999 999 s
up to 8 characters
read only
>RS232-settings
baud rate:
data bit:
stop bit:
parity:
handshake:
RS control:
Settings of RS232 interface.
Baud rate.
Data bit.
Stop bit.
Parity.
Handshake.
Receiving of commands via RS.
300,600,1200, 2400, 4800,
9600
7,
8
1
, 2
none
, odd, even
HWs
, HWf, SWchar, SWline, none
ON
, OFF
>common variables
Common variables C30...C39.
>prep.dosing elements
power ON prep:
report:
dos.element:
warn.interv.
dos.drive:
Preparation of titration burets.
Warning after power ON.
Report.
Selection of dosing unit.
Warning interval.
Selection of type of dosing unit.
ON,
OFF
ON,
OFF
internal D0
, external D1/D2
5...9999 min,
OFF
Dosimat
, Dosino
volume DX
cycle DX
dos.rate.DX
fill rate.DX
Parameters for the internal dosing unit
and for 685 Dosimat
:
Volume.
Number of cycles.
Dosing rate.
Filling rate.
0...
3.5
...99 999.99 ml
1...
2
...9
0.01...150 ml/min,
max.
0.01...150 ml/min,
max.
output:
len.dos.tub.
diam.dos.tub.
len.asp.tub.
diam.asp.tub.
dos.rate DX
fill rate DX
Parameters for the 700 Dosino
:
Location, where liquid is expelled.
Length of the dosing tube.
Diameter of the dosing tube.
Length of the aspiration tube.
Diameter of the aspiration tube.
Dosing rate.
Filling rate.
tip
, flask
1.0...
40.0
...999.9 cm
0.1...
2.0
...9.9 mm
1.0...
25.0
...999.9 cm
0.1...
2.0
...9.9 mm
0.01...150 ml/min,
max.
0.01...150 ml/min,
max.