2.10
Calculations
736 GP Titrino: Titration, dosing and measuring modes
24
Meaning
Meaning of
of the
the calculation variables
calculation variables
C variables are
•
determined by the method (C24...C27, C4X, C5X, C6X,
C7X, C8X)
•
entered as sample specific data (C00, C21...C23)
•
given as fixed constants in the methods (C01...C19)
Variable Meaning
C00
C01...C19
C21...C23
C24, C25
C26, C27
C30...C39
C40
C41
C42
C43
C44
C45
C46
C47
C51...C59
C61...C69
C70...C79
C80
C81...C89
Sample size, key <smpl data>.
Method specific calculation values, such as molecular mass, factors, key <C-
fmla>.
Sample specific calculation values, such as dilution factors, key <smpl data>.
Variables for storing determination results in the silo memory.
Means from silo calculations.
Common variables, e.g. for titer.
Initial measured value of the sample.
End volume.
Determination time.
Volume drift for KFT and SET with conditioning.
Temperature.
Dispensed start volume.
Asymmetry pH (pH calibration).
Electrode slope (pH calibration).
Fix EP for DET and MET or fix V for STAT.
pK/HNP values for DET and MET or fix times for STAT.
Temporary variables for calculations in TIP.
Mean rate for STAT: Overall rate, evaluated over all points of the measuring list.
Rates, evaluated in preset time windows for STAT.
Sample
Sample data,
data, key <
key <smpl
smpl data>
data>
•
Sample identifications or sample specific calculation val-
ues C21...C23
•
Sample size C00
For working with the silo memory, see page 27.
Display
Meaning
Input range
id#1 or C21
id#2 or C22
id#3 or C23
smpl size
smpl unit:
Inquiries with silo = OFF (LED ”silo” is
OFF):
Sample identification. Can be used as
sample specific calculation values.
Sample size
Unit of sample size
up to 8 ASCII characters
-999 999...
1
...999 999
g
, mg, ml, ul, pc, no unit or up to
5 ASCII characters
}