7 Troubleshooting
■■■■■■■■
67
3
Connect the instrument to the software as described in
Connecting the software with the instrument (page 22)
4
Load the example method
Dummy cell test.detp
5
Click on or
Menu bar
▶
Devices
▶
Start
to start the dummy cell
test.
Example of a dummy cell test measurement
Criteria to assess the dummy cell measurement
■
The curve must be a straight line, without spikes or steps.
■
At -1.0 V the current must be –0.89 - –1.09 µA.
■
At +1.0 V the current must be 0.89 - 1.09 µA.
To read out the exact current value at a specific potential, point the cursor
on the curve. Below the cursor the software will display the name of the
curve as well as potential and current at the specific point, e.g.
LSV_0001_sample_rep_01: (-1.0, -0.996)
. The first value in the brack-
ets is the potential, the second value the current.
NOTICE
If one of these criteria is not fulfilled, contact the local Metrohm Ser-
vice.