7 Troubleshooting
■■■■■■■■
47
Problem
Cause
Remedy
Technical failure.
Disconnect the USB cable from the potentio-
stat and try to switch it on again. If the green
status LED indicates that the potentiostat is on,
reconnect the USB cable.
Potentiostat - the
potentiostat
switches off during
a determination.
Battery level too low.
Connect the potentiostat to the power grid
using the USB power supply unit
. Charge the battery for at least
1 hour before switching on the potentiostat
again.
SPE - liquid is found
close to the connec-
tion socket for the
SPE.
The nut is not tightened
enough.
Fasten the nut until the slot in the silicone seal
completely seals the SPE.
The silicone seal is defec-
tive.
The silicone seal is a consumable. The sharp-
edged SPEs can cut into the silicone. It is rec-
ommended to replace the silicone seal every
25 electrodes.
The supporting ring is bent.
The supporting ring deforms over time.
Replace the supporting ring when it is
deformed and therefore does not equally dis-
tribute the pressure anymore.
SPE - the SPE cannot
be inserted into the
holder.
The nut is already too tight
and seals the slot in the sil-
icone seal.
Unfasten the nut to release pressure from the
silicone seal which opens the slot for the SPE.
The supporting ring turned
out of position.
Unfasten the nut and turn the supporting ring
so that the slot in the supporting ring is con-
gruent with the slot in the silicone seal. Also
make sure that the planar side of the support-
ing ring is attached to the silicone seal.
Stirrer - the stirrer
does not rotate.
The electrode cable is not
correctly connected.
Check that the plugs are connected with the
correct orientation
.
The electrode cable is
loose.
Make sure that the screws of the cable con-
nections are securely tightened.
The stirring and/or waiting
time in the method is 0.
The stirrer only runs during stirring and poten-
tiostatic pretreatment. Load the method and
set the correct time for
Stirring time
,
Wait-
ing time 1
and
Waiting time 2
.