7.3 Application problems
■■■■■■■■■■■■■■■■■■■■■■
56
■■■■■■■■
Problem
Cause
Remedy
sider the linear working range of the method
The observed peak is not
from the analyte.
■
Adapt the evaluation so that the correct
peak is evaluated (see
line parameters for all curves (page 30)
Adapting the base points for an individual
curve (page 31)
).
■
Identify the interference and remove it
from the measuring solution.
Peak evaluation -
the peaks are not or
not correctly evalu-
ated.
Peak shifted.
There are various reasons why a peak can shift
(see
Peak position - the peak in the sample as
well as for the standard addition has shifted.
).
Identify the problem and eliminate it if possi-
ble.
Wrong characteristic
potential.
Adapt the method parameter
Characteristic
potential
on the tab
Evaluation
in accor-
dance with the peak position
the peak recognition", page 29)
Wrong evaluation parame-
ters.
The method parameters
Tolerance
,
Min.
width
,
Max. width
and
Min. measured
quantity
on the tab
Evaluation
are thresh-
olds for the peak acceptance. Check these
parameters and adapt them in accordance
with the peak shape.
Substance evaluation not
active.
Make sure that the checkmark in the column
Active
, table
Substances
on the tab
Evalua-
tion
, is set.
Original curves are dis-
played.
Make sure that the option
Show original
curves
under
Main menu
▶
View
is not
checked.
Peak height - the
replicative measure-
ments differ signifi-
cantly.
The electrode is not in a
good condition.
Clean the electrode as recommended in the
application documentation and repeat the
determination. If the situation does not
improve, replace the electrode by a new one.
The curves are generally
very noisy.
See
Curve shape - the curves are very noisy.