2.8 Suppressor module
792 Basic IC
39
4 Connect aspirating tubing for H
2
O
x
Prepare aspirating tubing
68
68
68
68
: Cut a piece of the 6.1803.020
PTFE tubing to the required length (normally ca. 120 cm).
x
Insert one end of the aspirating tubing
68
68
68
68
into the rinsing
solution storage bottle and tighten it so that the tubing is
firmly held.
x
Insert the free end of aspirating tubing
68
68
68
68
into one of the
openings
5555
of the 792 Basic IC (
see Fig. 2
) and pull it suffi-
ciently far into the interior.
x
Mount a coupling
70
70
70
70
(6.2744.030) to the inlet end of the front
pump tubing
72
72
72
72
.
x
Mount a 6.2744.010 Compression fitting at the end of the
aspirating tubing
68
68
68
68
and screw this compression fitting on to
the coupling
70
70
70
70
(see
Fig. 14
).
2.8.4
Connection of the suppressor module
The three inlets and outlets numbered 1...3 on the suppressor module
39
39
39
39
each have 2 permanently mounted PTFE capillaries, which must be
connected as follows (see
Fig. 14
and
Fig.
15
):
1 Inlet capillary for eluent
x
Screw inlet capillary
75
75
75
75
marked with "Eluent" of suppressor
module
39
39
39
39
to outlet end of separating column
66
66
66
66
using a
6.2744.010 Compression fitting.
2 Outlet capillary for eluent
x
Screw outlet capillary
76
76
76
76
marked with "Detector" of suppres-
sor module
39
39
39
39
to coupling
26
26
26
26
using a 6.2744.010 Compres-
sion fitting.
x
Screw inlet capillary
37
37
37
37
of detector block
38
38
38
38
to other end of
coupling
26
26
26
26
.
3 Inlet capillary for H
2
SO
4
x
Attach inlet capillary
78
78
78
78
marked with "H
2
SO
4
" of suppressor
module
39
39
39
39
using a 6.2744.010 Compression fitting to the filter
unit PEEK
28
28
28
28
connected to the rear pump tubing
71
71
71
71
.
4 Outlet capillary for H
2
SO
4
x
Pull outlet capillary
80
80
80
80
marked with "Waste" of the suppressor
module
39
39
39
39
from below through one of the openings
6666
out of
the inner compartment of the 792 Basic IC.
x
Lead outlet capillary
80
80
80
80
to a sufficiently large waste container
and fix it in place.