5 Notes – Maintenance – Faults
792 Basic IC
148
x
To install a new piston seal
89
89
89
89
the special tool
94
94
94
94
is used.
x
First the new seal is inserted firmly in the recess of tool
94
94
94
94
by
hand (see
Fig. 17B
). The seal spring must be located on the
outside.
x
The tool
94
94
94
94
together with the seal is then inserted in the pump
head
34
34
34
34
and the seal pressed into the pump head recess
with the aid of tool
93
93
93
93
(see
Fig. 17C
).
The seal surface in the pump head
34
34
34
34
must not be damaged (avoid
contact with tool)!
5 Cleaning/replacement of inlet valve
90
90
90
90
and outlet
valve
91
91
91
91
x
Contaminated or blocked valves are cleaned by rinsing with
dist. water, RBS solution or acetone. The rinsing effect can be
reinforced by brief treatment in an ultrasonic bath (max. 20 s;
if longer the sapphire sphere of the valve can be damaged).
x
If this does not have the desired effect, the valves can be
disassembled as shown in
Fig. 18
. The valve components are
pushed out with the aid of a syringe needle inserted through
the upper opening in the valve housing
95
95
95
95
. The individual
components are rinsed with dist. water and/or acetone, and
the sapphire sphere cleaned with a paper towel. The valve is
then reassembled in accordance with
Fig. 18
. The compo-
nents of the inlet and outlet valves are identical, they are dis-
tinguished only by the positioning of the sapphire sleeve
98
98
98
98
and the ceramic holder
100
100
100
100
(see
Fig. 18
).
x
Valves that fail to function faultlessly after such cleaning must
be replaced.
x
In the reinstallation of the inlet valve
90
90
90
90
or the outlet valve
91
91
91
91
on no account must the two outwardly identical valves be in-
terchanged. To determine which valve is which, note that the
liquid flows through the pump head from the bottom up. The
flow direction of the valves can be checked simply by blowing
through the clean valve. Both valves are installed with the
black face in the direction of the pump head (see
Fig. 18
).
If by mistake an inlet valve
90
90
90
90
is installed instead of the outlet valve
91
91
91
91
,
an extreme pressure buildup occurs within the working cylinder, which
is not detected by the pressure transducer and will destroy the piston
seal