4.4 Methods
792 Basic IC
93
Time
Start time for integration event in min.
Event
List box with integration events to be selected
(see below).
Integration events
Disable detection
Set this mode to stop detection of new peaks. If
a peak is available at the moment of the event it
is either terminated (down slope peaks) or re-
jected (upslope peaks).
Enable detection
Clear
Disable detection
mode.
Set peak start
Force the beginning of a new peak. If a peak is
available at the moment of the event it is either
rejected (upslope) or terminated (down slope).
Set peak end
Force peak end at the time of the event. Up-
slope peaks are rejected (except those born by
Set peak start
event), down slope peaks are ter-
minated.
Enable valley-to-valley
Disable valley-to-valley
Enable valley-to-valley
Disable perpendicular drop peak separation. All
peaks are considered to be baseline-separa-
ted. The bottom of the valley becomes the
baseline point.
Disable valley-to-valley
Enable perpendicular drop peak separation
(default setting).