
4 Operation
792 Basic IC
104
Base
Base for calculations (
Area
or
Height
) in quantifi-
cation and calibration procedures. Indicates,
which parameter is to be used as a peak re-
sponse.
Reference channel
Channel which is used to measure area or
height (read-only).
Formula
Formula for calibration function. There are six
possible calibration dependencies for linear
and nonlinear calibration curves.
Linear through zero
function Y =
k
1
X
Linear
function Y =
k
1
X +
k
0
Quadratic through zero
function Y =
k
1
X +
k
2
X
2
Quadratic
function Y =
k
1
X +
k
2
X
2
+
k
0
Cubic through zero
function Y =
k
1
X +
k
2
X
2
+
k
3
X
3
Cubic
function Y =
k
1
X +
k
2
X
2
+
k
3
X
3
+
k
0
Statistical weight
Weighting parameter used to weight calibration
points deviations on calculating coefficients of
the calibration formula. Apart from the uniform
weighting
1
the values
1/x
,
1/x
2
,
1/y
and
1/y
2
are
also possible.
Standard component
Name of the standard component in the case
of
Internal standard calibration
or
Tabulated calibra-
tion
method (details see on-line help).
Concentration of standard
Concentration of the standard component in
the case of
Internal standard calibration
or
Tabu-
lated calibration
method (details see on-line
help).
Update calibration
792 BASIC IC / Method / Calibration / Update
Selection of this menu item opens the
Update
window.
Retention time
Replace the expected retention times of the
components with the retention times of the cor-
responding peaks obtained in the current run.
Содержание 2.792.0020 Basic IC
Страница 53: ...2 Installation 792 Basic IC 44...
Страница 197: ...6 Appendix 792 Basic IC 188...