2 Installation
792 Basic IC
16
2.3.5
Filter unit PEEK
One
6.2821.100 Filter unit PEEK
(see
Fig. 7
) is already installed be-
tween the high-pressure pump and the injection valve
22
22
22
22
at the 792 Ba-
sic IC. This filter unit serves to avoid contamination by abrasive particles
of the piston seals.
The two other filter units PEEK supplied with the 792 Basic IC are in-
stalled between the pump tubings of the peristaltic pump and the inlet
capillaries for regeneration and rinsing solution (see
section 2.8.2
).
These filter units serve to protect the suppressor module from foreign
particles and bacterial growth.
The Filter unit PEEK
28
28
28
28
consists of the housing
50
50
50
50
and the two connec-
tors
49
49
49
49
(with filter) and
51
51
51
51
(without filter) screwed into the housing
50
50
50
50
.
For the connection of capillaries
48
48
48
48
, PEEK compression fittings
46
46
46
46
(6.2744.010) or
47
47
47
47
(6.2744.070) must be used. New connectors
49
49
49
49
with
filter are available as an option with the ordering number 6.2821.110
(set of 10).
For the connection of the filter unit, please note the flow direction
arrow printed on the housing.
48
48
48
48
46
46
46
46
49
49
49
49
50
50
50
50
51
51
51
51
46
46
46
46
48
48
48
48
28
28
28
28
Fig. 7: 6.2821.100 Filter unit PEEK
28
28
28
28
Filter unit PEEK (6.2821.100)
49
49
49
49
Connector with filter (6.2821.110)
Part of 6.2821.100 Filter unit
46
46
46
46
Compression fitting (6.2744.010)
50
50
50
50
Housing for filter unit
Part of 6.2821.100 Filter unit
48
48
48
48
Capillary
6.1831.010 PEEK capillary or
6.1822.010 PTFE microcapillary
51
51
51
51
Connector without filter
Part of 6.2821.100 Filter unit
Содержание 2.792.0020 Basic IC
Страница 53: ...2 Installation 792 Basic IC 44...
Страница 197: ...6 Appendix 792 Basic IC 188...