5.2 Maintenance and servicing
792 Basic IC
153
105
105
105
105
104
104
104
104
103
103
103
103
102
102
102
102
Fig. 19: Assembling the suppressor
102
102
102
102
Screw nut
104
104
104
104
Suppressor rotor (6.2832.000)
103
103
103
103
Connection piece (6.2832.010)
with input and output leads
105
105
105
105
Suppressor holder
5.2.9
Replacing the suppressor
The suppressor in the suppressor block may have to be replaced in the
following cases:
x
Irremediable loss of suppressor capacity (reduced
phosphate sensitivity and/or strong rise in baseline)
x
Irremediable blockage of the suppressor (the suppres-
sor can no longer deliver solutions)
Both the 6.2832.000 Suppressor rotor and the 6.2832.010 Connection
piece with the input and output leads can be replaced. To replace these
components proceed as follows (see
Fig. 19
):
1 Disconnect suppressor from IC system
x
Disconnect all input and output leads of the suppressor from
IC system and peristaltic pump.
2 Dismantle suppressor
x
Unscrew nut
102
102
102
102
from suppressor holder
105
105
105
105
.
x
Pull out connection piece
103
103
103
103
and suppressor rotor
104
104
104
104
from
suppressor holder
105
105
105
105
(the connection piece and the rotor
normally stick together).
x
Loosen connection piece
103
103
103
103
from suppressor rotor
104
104
104
104
.
Содержание 2.792.0020 Basic IC
Страница 53: ...2 Installation 792 Basic IC 44...