12.8 The Limits and their Meanings
726 Titroprocessor, Short Introduction
50
50
Press the [Sample silo] softkey on the GLP main page to enter the sample
data. The sample sizes should scatter randomly within wide ranges. The
sample silo can be saved and restored later.
The validation can be started by pressing the [Start valid.] softkey. You will
be asked then to answer the following questions:
Do you really want to overwrite actual method [y]
Do you really want to reset all statistics data [y]
Do you want to set statistics "on" [y]
Do you want to set silo "on" [y]
Do you want to set auto start "on" [y]
Did you set up all necessary silo data [y]
Do you change samples manually [n]
(*)
Accept all previous settings [y]
If you confirm all default settings by pressing <enter> you can proceed
with the validation after entering your user name.
Validation from xx-xx-xx done by user: [ ]
Explanation:
It is recommended to save all actual data (method and determinations) be-
fore performing a validation. Correct statistical calculations require the
clearing of the working memory.
Validations should be done with auto start and silo „on“. If a sample
changer is connected, you are free to choose to use the changer or to
change the samples manually, see (*).
After finishing the validation series a validation report will be printed auto-
matically and the results may be viewed on the GLP page (<GLP> key).