9.1 Titration and Measuring Modes
726 Titroprocessor, Short Introduction
28
28
Measuring Modes:
MEAS_PH
Measurement,
measured value pH
MEAS_U
"
measured value mV
MEAS_UPOL
"
with polarization voltage
MEAS_IPOL
"
with polarization current
MEAS_T
"
measured value °C
CAL
Calibration,
measured value pH
The parameters of a titration or measuring mode can be set after pressing
the [* Param] softkey. Each mode has ist own parameter set.
Example:
ÚÄÄÄ Method : Chloride.mth ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
³ ³
³ t/s Command Parameters Parameters ³
³ ÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ³
³ 1 10 ADÚÄÄÄ DET_U* Parameters ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿ ³
³ ³ ³ ³
³ ³ Start conditions ³ ³
³ 2 DE³ Volume off Meas. val off mV ³ ³
³ ³ Dos. rate max mL/min Pause 0 s ³ ³
³ ³ ³ ³
³ ³ Titration conditions ³ ³
³ 3 EN³ Adjustment medium Time recording on ³ ³
³ ³ Temperature 25.0 øC ³ ³
³ ³ ³ ³
³ ³ Stop conditions ³ ³
³ ³ Volume absolute Meas. val off mV ³ ³
³ ³ Size 99.99 mL EP# 9 ³ ³
³ ³ Fill. rate max mL/min Stop time off s ³ ³
³ ³ Mpt. overflow on ³ ³
³ ³ ³ ³
³ ³ ³ ³
³ ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ Quit ÄÙ ³
³ ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ Quit ÄÙ
ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿
³ ³ ³ ³ ³Recogn.³Display³Custom ³ ³ ³ ³ ³
³ ³ ³ ³ ³ ³ ³adjust ³ ³ ³ ³ ³
ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ
The titration modes are structured into ‘start conditions’, ‘titration conditions’
and ‘stop conditions’.
‘Start conditions’
Actions or conditions to be executed before the
regular, controlled titration, e.g.
Start volume
or
Pause
(after distribution of start volume).
‘Titration conditions’
Control parameters of the titration
e.g.
Adjustment [slow ³ medium ³ fast].
These
entries are sets of control parameters that may be
customized if necessary, see softkey [Custom ad-
just].
‘Stop conditions’
Settings to determine the termination of the titration,
e.g.
EP#
(stop when endpoint reached),
Stop volume
.
The conditioning modes SEC and KFC only show ‘Titration conditions’.
The calibration mode CAL* shows the buffer table where buffer types and
values can be entered.
The measuring mode MEAS shows ‘Measure conditions’ which allow to
perform multiple drift-controlled measurements to optain a measuring curve.
Note: Do not define an Equilibr. time greater than the cycle time.